In this post Standard Operating Procedure Cleaning, Operation and Calibration of Disintegration Test Apparatus in Pharmaceuticals is describe.
1.0 OBJECTIVE:
To lay down a procedure for Cleaning, Operation and Calibration of Disintegration Test Apparatus.
2.0 SCOPE:
This procedure is applicable to Cleaning, Operation and Calibration of Disintegration Test Apparatus used in Production Department.
3.0 RESPONSIBILITY:
Officer, Executive Production : Operation and Calibration Head Production : SOP Compliance
IPQA : Operation and Calibration
4.0 DEFINITION(S):
NA
5.0 PROCEDURE:
5.1 Cleaning:
5.1.1 Cleaning of basket and disc:
5.1.2 Clean the basket of the disintegration apparatus with 5-10 liters purified water use soft nylon brush if required.
5.1.3 Clean the body with a wet lint free duster.
5.1.4 Clean the basket and discs after every disintegration test.
5.1.5 Frequency: After every test or weekly which ever is earlier.
5.1.6 Cleaning of water bath:
5.1.6.1 Clean the water bath with 5-10 liters of purified water use soft nylon brush if required.
5.1.6.2 Clean the body with a wet lint free duster if not in use.
5.1.6.3 Frequency: Weekly
5.2 Operation:
5.2.1 Ensure the cleanliness of the instrument.
5.2.2 Switch ‘ON’ mains Switch.
5.2.3 Put the power switch ‘ON’ at the rear panel.
5.2.4 Set the conditions: (Be sure that water is filled upto the mark in the water bath).
5.2.5 Setting the temperature:
5.2.5.1 Press “SET” key the yellow LED starts glowing.
5.2.5.2 Press ‘TEMP’ key, the right digit of the previously set temperature on the temperature screen starts blinking.
5.2.5.3 Set the required temperature with pressing Up/Down keys. After the required setting, press “DISP SEL” key to change the position of blinking digit. After complete setting press “ENTER” key to store the temperature set in instrument’s memory.
5.2.6 Setting the timer:
5.2.6.1 Press “SET” key the yellow LED on the right side starts glowing.
5.2.6.2 Press “TIMER” key, the right digit of the previously set time starts blinking.
5.2.6.3 Enter the required time in minutes with pressing up / down keys. After the required setting, press “DISP SEL” key to change the position of the blinking digit. After complete settings press “ENTER” key to store the time set in instrument’s memory.
5.2.6.4 Press the “TEMP” key to start the heating of water in water bath. The red “LED” starts glowing.
5.2.6.5 Take two 1000 ml beakers and fill them with purified water/specified media upto the ¾áµ—ʰ level. Place them in the instrument.
5.2.6.6 Take the temperature of the water bath / beaker solution by keeping the external probe in that and press “PROBE SEL” key. The LED of “EXT.” position starts glowing and the temperature of that area will be displayed.
5.2.6.7 After attaining the set temperature, place the tablets/capsules in the basket (use discs, if required), fix them on the clamps and start the test by pressing “START/STOP” key. The instrument will automatically stop after completing the set time.
5.2.6.8 To halt the test, press “TIMER” key. The green LED of “TIMER” key start flashing and the time stop increasing. To continue the test press “TIMER” key once more.
5.2.6.9 To stop the test whenever required press “START/STOP” key, the test will stop.
5.2.6.10 Clean the baskets, discs and beakers for next use.
5.2.6.11 Put “OFF” the instrument from the rear panel and switch “OFF” the mains.
5.3 Calibration:
5.3.1 Operate the Instrument as per the step no. 5.2.
5.3.2 Start the machine and count the number of cycles per minute and record the same in Annexure- I.