Ad Code

HPLC Column Washing Procedure (SOP) for Pharmaceutical QC Laboratories

Step-by-step HPLC column washing procedure (SOP) for C18, C8, Phenyl, Protein Pack, and L17 columns used in pharmaceutical analysis. Ensure column performance, longevity, and compliance with proper HPLC column washing techniques.

HPLC Column Washing Procedure (SOP)

Standard Operating Procedure (SOP) for HPLC Column Washing

1.0 Objective

To describe the standard procedure for HPLC column washing to ensure optimal column performance, extended column life, and reliable analytical results in pharmaceutical product analysis.


2.0 Scope

This SOP applies to the washing of HPLC columns used in the Quality Control laboratory for the analysis of pharmaceutical products.


3.0 Responsibility

Officer/Executive – Quality Control


4.0 Accountability

Manager – Quality Control


5.0 Procedure for HPLC Column Washing

Proper HPLC column washing is essential to remove residual buffers, salts, and analytes that may affect chromatographic performance. The washing procedure varies depending on the column chemistry.


5.1 HPLC Column Washing for C18, C8, C6, Phenyl, CN, and Amine Columns

Follow these steps carefully:

Step 1: Flush with Water

  • Replace the mobile phase with filtered distilled water.
  • Set the flow rate to 1.0 mL/min.
  • Flush for 30 minutes.
  • Ensure system pressure does not exceed 400 bar.

Step 2: Flush with 80% Acetonitrile (ACN)

  • Change the mobile phase to HPLC-grade 80% Acetonitrile.
  • Maintain a flow rate of 1.0 mL/min.
  • Wash for 30 minutes.

Step 3: Storage

  • Ensure the column is filled with 80% ACN.
  • Disconnect the column from the HPLC system.
  • Store in an appropriate, clean, and dry location.

This method ensures effective removal of aqueous buffers and hydrophobic contaminants from reversed-phase columns.


5.2 HPLC Column Washing for Protein Pack Columns

Protein-based columns require a slightly different cleaning approach.

Step 1: Prepare Washing Solution

  • Prepare a 1:10 mixture of Isopropyl Alcohol and HPLC water.

Step 2: Flush the Column

  • Set flow rate to 1.0 mL/min.
  • Wash for 30 minutes.
  • Ensure system pressure does not exceed 400 bar.

Step 3: Storage

  • Confirm the column contains the washing solution.
  • Disconnect and store properly.

This procedure prevents protein precipitation and maintains column efficiency.


5.3 HPLC Column Washing for L17 Columns

L17 columns require controlled acidic flushing for proper maintenance.

Step 1: Acidic Flush

  • Prepare a mixture of 0.01 N H₂SO₄ : ACN (85:15).
  • Set flow rate to 0.6 mL/min (do not exceed this rate).
  • Flush for 2 hours.

Step 2: Water Flush

  • Change mobile phase to water.
  • Maintain flow rate at 0.6 mL/min.
  • Wash for 2 hours.

Step 3: Storage

  • Ensure the column contains water.
  • Disconnect and store appropriately.

Following this extended HPLC column washing process ensures complete removal of acidic residues and maintains column integrity.


6.0 Abbreviations

  • SOP – Standard Operating Procedure
  • ACN – Acetonitrile
  • HPLC – High Performance Liquid Chromatography


Why Proper HPLC Column Washing Is Important

Routine and correct HPLC column washing:

  • Prevents salt precipitation
  • Removes buffer residues
  • Reduces backpressure issues
  • Extends column life
  • Maintains reproducibility
  • Ensures regulatory compliance

In pharmaceutical QC laboratories, improper column washing can lead to inconsistent results, costly downtime, and column damage.


Frequently Asked Questions (FAQ)

1. Why is HPLC column washing necessary?

HPLC column washing removes residual buffers, salts, and analytes that can cause high backpressure, peak tailing, and reduced column efficiency.

2. How often should HPLC columns be washed?

Columns should be washed after completing analysis involving buffers, salts, or strongly retained compounds, and before long-term storage.

3. Why is 80% Acetonitrile used for washing C18 columns?

80% ACN effectively removes hydrophobic compounds and prevents microbial growth during storage.

4. What happens if an HPLC column is not properly washed?

Improper washing can result in increased pressure, poor peak shape, reduced resolution, and permanent column damage.

5. Can the same washing procedure be used for all HPLC columns?

No. Different column chemistries (e.g., C18, Protein Pack, L17) require specific washing procedures to maintain performance and longevity.