In pharmaceutical chemistry, precision is everything. One of the most fundamental tools used to ensure the identity and purity of drug substances is UV-Visible Absorption Spectrophotometry. This technique measures how a chemical substance in solution absorbs light across specific wavelengths, providing a unique "optical fingerprint."
This article explores the principles, apparatus requirements, and essential calibration controls for UV-Vis spectrophotometry as per pharmacopoeial standards.
What is UV-Visible Spectrophotometry?
UV-Vis spectrophotometry involves the measurement of the absorption of monochromatic (single-wavelength) radiation by a solution. The spectrum is divided into two primary ranges:
- Ultraviolet (UV) Range: 185 nm to 380 nm.
- Visible Range: 380 nm to 780 nm.
The Concept of Absorbance (A)
Absorbance is defined as the logarithm (base 10) of the reciprocal of transmittance (T). In simpler terms, it measures how much light is "trapped" by the molecules in the solution.
Specific Absorbance: A(1%, 1 cm)
To standardize calculations, pharmacopoeias use Specific Absorbance. This refers to the absorbance of a 1% w/v solution measured in a 1 cm cell. The formula used is:
Where:
- A = Measured absorbance.
- c = Concentration of the substance in percentage w/v.
- l = Thickness of the absorbing layer (path length) in cm.
The Apparatus: How a Spectrophotometer Works
A standard spectrophotometer consists of an optical system that produces monochromatic light between 200 nm and 800 nm.
Key Requirements:
- Cuvettes (Cells): Two empty cells (one for the sample and one for the reference) must have identical spectral characteristics.
- Reference Liquid: Usually, the same solvent used to dissolve the sample.
- Temperature: Unless otherwise stated, measurements are typically performed between 24 C and 26 C.
Critical Quality Controls (Calibration)
To ensure valid results, the spectrophotometer must undergo regular performance verification. There are four critical checks required:
1. Control of Wavelengths
The wavelength scale must be accurate to ensure the drug is being measured at its true maximum absorption.
- Reference Standards: Holmium perchlorate solution, hydrogen/deuterium lamps, or mercury vapor lines.
- Tolerance: ± 1nm (200-400 nm) and 3 nm (400-600 nm).
2. Control of Absorbance
This verifies the photometric accuracy of the detector.
- Method: A solution of Potassium Dichromate (dried at 130 C) in 0.005 M sulfuric acid is used.
- Checkpoints: Common wavelengths include 235 nm, 257 nm, 313 nm, 350 nm, and 430 nm.
- Tolerance: ± 0.01 absorbance units.
3. Limit of Stray Light
Stray light is "extra" light that hits the detector but does not pass through the sample, causing false readings.
- Test: A 1.2% w/v solution of Potassium Chloride (KCl).
- Limit: The absorbance should be greater than 2.0 at approximately 200 nm when compared to water.
4. Resolution Power
Resolution ensures the instrument can distinguish between two closely spaced peaks.
- Test: Recording the spectrum of a 0.02% v/v solution of Toluene in Hexane. The ratio of the maximum absorbance to the minimum absorbance at specific peaks is used to determine resolution.
