Ad Code

SOP for Quality Control (QC) Laboratory Instruments

SOP for Quality Control (QC) Laboratory Instruments

 Precision is the backbone of the QC Lab. When high-performance instruments like HPLCs, GCs, or Spectrophotometers fail, the following steps are mandatory:

  1. Cease Operations: Stop all testing on the instrument immediately upon calibration failure.

  2. Tagging: Place an "UNDER MAINTENANCE" sign on the system.

  3. Expert Intervention: For malfunctions or "Out of Calibration" results, contact the original manufacturer or authorized supplier to request a Service Engineer.

  4. Rectification: The service engineer must investigate the root cause, perform the necessary repairs, and provide a detailed service report.

  5. Re-Validation/Calibration: After maintenance, the lab must perform either a partial or complete calibration to verify the instrument's performance.

  6. Documentation: File the service report (detailing the cause of failure and actions taken) alongside the new calibration report in the instrument's history file.

  7. Impact Assessment: The investigation must determine if the failure could have affected the results of samples tested since the last successful calibration.

  8. QA Evaluation: The Head of Quality Assurance (QA) will evaluate the need to re-analyze previous samples. This ensures that no "false pass" results were generated due to the faulty equipment.


Summary of Comparison: Engineering vs. QC Lab

FeatureGeneral EquipmentQC Laboratory
Primary ResponderInternal Engineering DeptExternal Service Engineer/Manufacturer
DocumentationIncident ReportService Report + Calibration Report
Quality CheckProduct Impact AssessmentSample Re-analysis (QA Lead)
ReplacementChange Control (if non-identical)Performance Qualification (PQ) required