Precision is the backbone of the QC Lab. When high-performance instruments like HPLCs, GCs, or Spectrophotometers fail, the following steps are mandatory:
Cease Operations: Stop all testing on the instrument immediately upon calibration failure.
Tagging: Place an "UNDER MAINTENANCE" sign on the system.
Expert Intervention: For malfunctions or "Out of Calibration" results, contact the original manufacturer or authorized supplier to request a Service Engineer.
Rectification: The service engineer must investigate the root cause, perform the necessary repairs, and provide a detailed service report.
Re-Validation/Calibration: After maintenance, the lab must perform either a partial or complete calibration to verify the instrument's performance.
Documentation: File the service report (detailing the cause of failure and actions taken) alongside the new calibration report in the instrument's history file.
Impact Assessment: The investigation must determine if the failure could have affected the results of samples tested since the last successful calibration.
QA Evaluation: The Head of Quality Assurance (QA) will evaluate the need to re-analyze previous samples. This ensures that no "false pass" results were generated due to the faulty equipment.
Summary of Comparison: Engineering vs. QC Lab
| Feature | General Equipment | QC Laboratory |
| Primary Responder | Internal Engineering Dept | External Service Engineer/Manufacturer |
| Documentation | Incident Report | Service Report + Calibration Report |
| Quality Check | Product Impact Assessment | Sample Re-analysis (QA Lead) |
| Replacement | Change Control (if non-identical) | Performance Qualification (PQ) required |
%20Laboratory%20Instruments.webp)