Preserving Sterility: Validating the Hold Time of Sterile Glassware and Accessories

sterility of glassware and accessories


In pharmaceutical Quality Control (QC) laboratories, especially in the Microbiology section, maintaining the sterility of glassware and accessories is absolutely essential. Equipment like media bottles, funnels, and sampling tools must remain free of microorganisms from the moment they are sterilized until they are used for testing.

Sterilized Hold Time Validation is the documented process that confirms that these materials retain their sterile state for a specified period when stored under controlled conditions. This protocol aims to scientifically establish a safe Hold Time for common lab items after they are processed in an autoclave.


🎯 Objective and Scope

Objective: To provide documented evidence, through scientific data, that no microbial presence is detected on sterilized glassware and accessories when held up to a specified time (e.g., 120 hours / 5 days) after sterilization, using a microbial detection method (Sterility Test).

Scope: This protocol specifically evaluates the hold time of various sterilized materials—including screw-capped bottles, Miliflex funnels, Stainless Steel (SS) frits, and tips—following their sterilization in an autoclave.

👥 Responsibility and Accountability

The success of this microbial validation relies on defined roles and cross-functional oversight:

  • Quality Control (QC): Responsible for the preparation, review, and execution of the protocol, including all sampling, testing, and initial documentation.
  • Quality Assurance (QA): Responsible for the review and final approval of the protocol, ensuring compliance with regulatory and internal standards.
  • Accountability: Final accountability rests with the Head of QA & QC.


🧪 Procedure: The Hold Time Study Execution

The study is designed to challenge the packaging and storage of the sterilized items. The pre-requisite is ensuring that environmental conditions (like those in a Laminar Air Flow unit) are maintained within established limits.

1. Sterilization and Storage

  1. Sterilization: The specified materials (bottles, funnels, frits, and tips) are prepared and subjected to sterilization according to the site's SOP for sterilization.
  2. Storage: After sterilization, the packaged materials are stored in a designated controlled environment, typically a Laminar Air Flow (LAF) unit, for the duration of the study.

2. Sampling and Testing Intervals

The Hold Time study tracks microbial integrity over a 120-hour period. Sampling is performed at regular intervals:

  • 0hr (Initial Sample, immediately after storage begins)
  • 24hr
  • 48hr
  • 72hr
  • 96hr
  • 120hr

3. Microbial Detection Method

At each interval, samples are collected aseptically and subjected to a sterility challenge using Soybean Casein Digest Medium (SCD medium)

Item TestedProcedure
Screw Capped BottlesAseptically pour SCD medium directly into the collected bottle.
Miliflex Cups/FunnelsAseptically filter 100mL of sterile water through the cup/funnel; transfer the membrane into SCD medium.
SS FritsAseptically transfer the SS frit into 100 mL of SCD medium.
TipsAseptically transfer the tip into 100 mL of SCD medium.

4. Incubation

All test samples are incubated for a total of 14 days under dual temperature conditions to promote the growth of any surviving or contaminating microorganisms:

  • 20-25C for 7 days
  • 30-35C for 7 days


📈 Acceptance Criteria and Conclusion

Acceptance Criteria

For the sterilization hold time to be established and verified, the results must demonstrate continued sterility:

  • Microbial Determination: No growth should be observed in the sterility test for samples collected at any time interval up to 120hours.

Conclusion

After the 120hr sample has completed its 14-day incubation period, the final Hold Time Study Summary Report is prepared. This report must clearly contain the discussion, data evaluation, and a definitive conclusion that formally establishes the maximum safe hold time (e.g., 120 hours) for the sterilized glassware and accessories in the microbiology lab.

This validation provides the critical assurance that all sterile components used in sensitive testing environments maintain their integrity, safeguarding the accuracy and compliance of pharmaceutical testing .