1.0 OBJECTIVE:
To
lay down the procedure for calibration of Tapped density tester.
2.0 SCOPE:
This
shall be applicable to the Tapped density tester of Quality Control Instrument
Lab.
3.0 RESPONSIBILITY:
3.1
Technical
Assistant/ Helper.
3.2
Q.C Analyst.
4.0 ACCOUNTABILITY:
4.1
Manager-Quality
Control.
5.0 MATERIAL AND
EQUIPMENT:
5.1
Tapped density
tester.
6.0 PROCEDURE:
6.1 Before
starting calibration, check cleanliness of instrument i.e. SOP.
6.2 Switch
“ON” the power supply.
6.3 Operate
the apparatus according to SOP for operation of tapped density tester i.e. SOP
No. BM/QCEO/SOP042-00.
6.4 Set
the instrument for 100, 250, 750 and 1250 taps.
6.5 Count
the tap value & enter observations.
6.6 Record
the result on the “Calibration Record”,
“Calibration Log Book” & paste the “Calibration
Status Sticker” on Tapped density tester.
7.0 REVISION LOG:
Revision No.
|
Effective Date
|
Reason
|
00
|
|
New
SOP
|
8.0 REFERENCES:
8.1
Not Applicable.
9.0 ANNEXURES:
Annexure 1: Calibration
of Tapped density tester.
Annexure:
1
Calibration
of Tapped density tester
Name
of instrument:
Instrument
serial No.:
Make
& Model:
Calibration
date:
Calibration
due date:
Weight of the sample=
___________g.
Remarks:
Satisfactory/ Not satisfactory.
|
10.0
ABBREVIATIONS:
Abbreviation
|
Expanded Form
|
SOP
|
Standard operating
procedure
|
&
|
And
|
No.
|
Number
|
Ltd.
|
Limited
|
QCI
|
Quality control
instrument lab
|
Sr.#
|
Serial number
|
Dept.
|
Department
|
Q.C
|
Quality control
|
g
|
Grams
|
Min
|
Minutes
|
%
|
Percentage
|
cm3
|
Cubic centimeter.
|
Vol
|
Volume
|