1.0 OBJECTIVE:
To
lay down the procedure for cleaning of Magnetic stirrer used for mixing of sample solution in Lab.
2.0 SCOPE:
This
SOP shall be applicable to the Magnetic
stirrer of installed, at Quality Control Instrument Lab.
3.0 RESPONSIBILITY:
3.1
Technical
Assistant/ Helper.
3.2
Q.C Analyst.
4.0 ACCOUNTABILITY:
4.1
Manager-Quality
Control.
5.0 MATERIAL AND
EQUIPMENT:
5.1
Magnetic stirrer with hot plate.
5.2
Dry cotton cloth.
5.3
Purified water.
6.0 PROCEDURE:
6.1
Switch “OFF” the
instrument after use.
6.2
Switch “OFF” the
mains and remove the plug from switch board.
6.3
Clean the
apparatus with clean and dry duster.
6.4
Wash the magnet
with purified water after every use and dry it by using dry cloth.
6.5
After completion affix
the “CLEANED” label and fill the “Cleaning Equipment Log Book”.
7.0 REVISION LOG:
Revision No.
|
Effective Date
|
Reason
|
00
|
|
New
SOP
|
8.0 REFERENCES:
8.1 Not Applicable.
9.0 ANNEXURES:
Annexure 1: Cleaning
record of magnetic stirrer.
Annexure: 1
Cleaning record of Magnetic
stirrer
CLEANING RECORD OF MAGNETIC
STIRRER
|
10.0
ABBREVIATIONS:
Abbreviation
|
Expanded Form
|
SOP
|
Standard operating
procedure
|
&
|
And
|
No.
|
Number
|
Ltd.
|
Limited
|
QCEC
|
Quality control Equipment
cleaning
|
Q.C
|
Quality control
|