1.0 OBJECTIVE:
To
lay down the operating procedure of Disintegration apparatus.
2.0 SCOPE:
This
shall be applicable to the Disintegration
apparatus of Quality Control Instrument Laboratory.
3.0 RESPONSIBILITY:
3.1
Technical
Assistant/ Helper.
3.2
Q.C Analyst.
4.0 ACCOUNTABILITY:
4.1
Manager-Quality
Control.
5.0 MATERIAL AND
EQUIPMENT:
5.1
Disintegration apparatus.
5.2
Purified water.
5.3
Sample (tablet/capsule).
6.0 PROCEDURE:
6.1
How
to operate the Machine
For easy and smooth operation follow
the following Steps
6.1.1
Connect the “Power
Plug” to the socket (220V, AC).
6.1.2
Turn ON the “Main
Button” which is located at the bottom of right side of the machine, Bulb in
the “Main Button” will illuminate.
NOTE:
If the bulb does not illuminate, check the main supply.
6.1.3
“LCD” (Liquid Crystal
Display) which is located on the front panel of the machine will be illuminated
and will show following display.
|
By pressing any key, following screen will be
displayed
MAIN
SV=
00:00:00 RPM 30
Temp 37oC
MOTOR= Off
|
6.1.4
To start the test
press 1 from the front panel, following screen will be displayed.
SV=
00:00:00 RPM 30
Temp 37oC
MOTOR= Off
|
6.1.5
Press “Start/Stop”
button on the front panel to start the test.
6.1.6
Press 8 (Heater)
button on the front panel to “Start/Stop” of Heater.
Note:
The temperature will be adjusted to 37oC automatically.
6.1.7
Press 6 (Cir)
button on the front panel to “Start/Stop” of PUMP.
6.1.8
Press 1 (Light)
button on the front panel to “Start/Stop” of Light.
6.1.9
For changes in
setting, press Esc following screen will be displayed.
The option → and ← is for
adjust Temperature , RPM and Time and Ent button for Save
Setting and then press Esc button
to enter Main Screen.
6.2 Precautions:
6.2.1
Do not switch on
the mains if water in the tank is not up to mark.
6.2.2
While placing and
removing the basket assembly does not apply excessive force.
7.0 REVISION LOG:
Revision No.
|
Effective Date
|
Reason
|
00
|
|
New
SOP
|
8.0 REFERENCES:
8.1
Not Applicable.
9.0 ANNEXURES:
Annexure
1 : Description
of front panel keys
Annexure: 1
Description of front
panel keys
DESCRIPTION OF FRONT
PANEL KEYS
1.
0
– 9 digit keys are for numeric entries.
2.
“
1 “ Key is for flash light, Press and hold “ 1 “ key the flash light will
illuminate.
3.
“6
“key is for water circulation.
4.
“8”
key is for heater.
5.
“Esc”
key is used to go back to previous menu.
6.
“Reset”
Key is used to reset the timer.
7.
“Start/Stop”
key is used to start or stop the oscillations.
8.
“Enter”
key is used to save the entries
9.
←
and
→ arrow keys are used to change the options
|
10.0
ABBREVIATIONS:
Abbreviation
|
Expanded Form
|
SOP
|
Standard operating
procedure
|
&
|
And
|
No.
|
Number
|
Ltd.
|
Limited
|
Q.C
|
Quality control
|
QCEO
|
Quality control
equipment operation
|
F
|
Format
|
oC
|
Degree centigrade
|
No comments: